• 美狮贵宾会

    产品资料

    Temperature/Climatic Stress Screening Chambers, Type 5k/min

    如果您对该产品感兴趣的话,可以
    产品名称: Temperature/Climatic Stress Screening Chambers, Type 5k/min
    产品型号:
    产品展商: Lab Companion
    产品文档: 无相关文档

    简单介绍

    Temperature/Climatic Stress Screening Chambers, Type 5k/min


    Temperature/Climatic Stress Screening Chambers, Type 5k/min  的详细介绍

    Temperature/Climatic Stress Screening Chambers, Type 5k/min

    The Lab Companion temperature/climatic stress screening test cabinets can achieve high change rates of up to 30 K/min. with conventional cooling technology thanks to amplified compressors and a strong fan capacity. By this means, CTS offers optimised output with low consumption figures and attractive prices. Thus, the limits of your test item can be determined more quickly. Faster change rates are optionally possible using nitrogen.



    Basic Equipment:

    • capacitive Multi-Touch-Display in the front
    • service-friendly capacitive humidity measurement system
    • defined keys for elementary functions of the installation such as start/stop, fixed-value/automatic operation, interruption, light, etc.
    • USB-Connection to save measuring data on USB Stick
    • Ethernet-Interface
    • 2 digital Out, to switch test specimen ON/OFF
    • adjustable software-temperature-limiter min./max.
    • shelf, stainless steel
    • entry port Ø 50 mm on the right side
    • single-hand operated door handle, lockable

    Your benefits:

    • low consumption data
    • minimum noise level
    • extremely high temperature change rates
    • user-friendly operation and programming
    • uniform operation due to identical Multi-Touch-Display for all chamber sizes
    • demineralized water system with level indicator
    • temperature measuring with Pt 100 in air inlet and air outlet
    • service-friendly construction
    • international standards are fulfilled (CE-Conformity)



    Temperature change rates:

    • 5 K/min, 10 K/min und 15 K/min

    Test space capacity in litres:

    • 100 / 200 / 350 / 600 / 1000 / 1500 / 2000

    Temperature range:

    • -40 °C / +180 °C 
      -70 °C / +180 °C

    Design:

    Control:
    Multi-Touch-Display with integrated program controller
    Test space:
    stainless steel grade 1.4301
    Refrigerants:
    environmental friendly refrigerants
    Humidity system (only type CS):
    capacitive measuring








    Options:

    • add. Entry ports Ø 50 mm, 80 mm und 125 mm
    • add. shelves
    • panoramic window, multiple glazed, sized 570 x 710 mm, < 200 ltr., incl. LED-light
    • hand-holes in the door
    • add. Pt 100 for temperature measuring on test specimen reversible to controller function
    • air dryer system
    • mobiles design
    • air/water cooled refrigeration unit
    • temperature protection for test specimen
    • CID software for programming and documentation
    • variable air speed in the test space
    • notch
    • Other interfaces: RS232, RS485 and USB

    Type 5K/min

    Type
    -70/100-5
    -40/200-5
    -70/200-5
    -40/350-5
    -70/350-5
    -40/600-5
    -70/600-5
    -40/1000-5
    -70/1000-5
    -40/1500-5
    -70/1500-5
    Test Space litres 100 200 350 600 1000 1500
    Test Space Dimensions in mm height, appr.
    width, appr.
    depth, appr.
    500
    500
    400
    750
    650
    400
    750
    650
    720
    850
    850
    830
    900
    1000
    1100
    900
    1000
    1600
    Overall Dimensions in mm height, appr.
    width, appr.
    depth, appr.
    1630
    770
    860
    1800
    920
    1345
    1800
    920
    1665
    1900
    1120
    1875
    1950
    1270
    2245
    1950
    1270
    2745
    Temperature Tests
    Temperature Range in °C TS -40/..., CS -40/...
    TS -70/..., CS -70/...

    -70/+180
    -40/+180 
    -70/+180
    -40/+180 
    -70/+180
    -40/+180 
    -70/+180
    -40/+180 
    -70/+180
    -40/+180 
    -70/+180
    Temp. Change Rate Cooling acc. IEC 60068-3-5 in K/min TS -40/..., CS -40/...
    TS -70/..., CS -70/...

    8,0
    5,0
    5,0
    5,0
    5,0
    6,0
    7,0
    5,0
    5,0
    5,0
    5,0
    Temp. Change Rate Heating acc. IEC 60068-3-5 in K/min TS -40/..., CS -40/...
    TS -70/..., CS -70/...

    6,0
    8,0
    8,0
    8,0
    8,0
    6,0
    6,0
    6,0
    6,0
    5,0
    5,0
    max. Thermal Load at +20°C in W TS -40/..., CS -40/...
    TS -70/..., CS -70/...

    3500
    3500
    3500
    3500
    3500
    4500
    5000
    5200
    5200
    5000
    5500
    Temperature Fluctuation in K ≤± 0,3 ≤± 0,3 ≤± 0,3 ≤± 0,3 ≤± 0,3 ≤± 0,3
    Climatic Tests
    Temperature Range in °C CS -40/..., CS -70/... +10 to +95 +10 to +95 +10 to +95 +10 to +95 +10 to +95 +10 to +95
    Humidity Range in % RH CS -40/..., CS -70/... 10 to 98 10 to 98 10 to 98 10 to 98 10 to 98 10 to 98
    Dew Point Range I in °C CS -40/..., CS -70/... +7 to +94 +7 to +94 +7 to +94 +7 to +94 +7 to +94 +7 to +94
    Dew Point Range II in °C CS -40/..., CS -70/... -10 to +7 °C -10 to +7 °C -10 to +7 °C -10 to +7 °C -10 to +7 °C -10 to +7 °C
    Humidity Fluctuation I in % RH CS -40/..., CS -70/... ≤± 1,5 ≤± 1,5 ≤± 1,5 ≤± 1,5 ≤± 1,5 ≤± 1,5
    Humidity Fluctuation II in % RH CS -40/..., CS -70/... ≤± 3 ≤± 3 ≤± 3 ≤± 3 ≤± 3 ≤± 3
    Temperature Fluctuation in K CS -40/..., CS -70/... ≤± 0,3 ≤± 0,3 ≤± 0,3 ≤± 0,3 ≤± 0,3 ≤± 0,3
    Nominal Voltage 400 V +6% -10%, 3/N, 50 Hz
    Nominal Power in kW TS -40/.., appr.
    CS -40/.., appr.
    TS -70/.., appr.
    CS -70/.., appr.


    6,5
    9,3
    10,4
    12,0
    12,0
    9,3
    10,4
    11,8
    11,8
    11,0
    12,1
    14,2
    14,2
    13,3
    13,2
    16,4
    16,4
    13,2
    13,2
    16,1
    16,1
    Refrigeration Unit air cooled air cooled air cooled water cooled water cooled water cooled
    Weight in kg TS -40/..., CS -40/..., appr.
    TS -70/..., CS -70/..., appr.

    300
    500
    550
    600
    600
    700
    750
    850
    900
    1000
    1100
    Noise level in dB(A) < 60 < 58 < 58 < 62 < 62 < 64

    Temperature Change Rate acc. IEC 60068-3-5:

    heating = Δ th / Δ Th

    cooling = Δ tk / Δ Tk

     

    Δ Th = time of heating
    Δ Tk = time of cooling
    Δ th = temperature difference, heating
    Δ tk = temperature difference, cooling

    All figures are average values which have been obtained at a temperature of +25° C, without test specimens, without thermal load and without options. We reserve the right to make alterations due to technical development.

    Other chamber sizes and options on request.




    产品留言
    标题
    联系人
    联系电话
    内容
    验证码
    点击换一张
    注:1.可以使用快捷键Alt+S或Ctrl+Enter发送信息!
    2.如有必要,请您留下您的详细联系方式!
    美狮贵宾会